Detector

The detector senses a physicochemical property of the analyte and provides a response which is amplified and converted into an electronic signal to produce a chromatogram. Most of the detectors used in GC were invented specifically for this technique, except for the thermal conductivity detector (TCD) and the mass spectrometer. In total, approximately 60 detectors have been used in GC. Detectors that exhibit an enhanced response to certain analyte types are known as "selective detectors".

During the last 10 years there had been an increasing use of GC in combination with mass spectrometry (MS). The mass spectrometer has become a standard detector that allows for lower detection limits and does not require the separation of all components present in the sample. Mass spectroscopy is one of the types of detection that provides the most information with only micrograms of sample. Qualitative identification of unknown compounds as well as quantitative analysis of samples is possible using GC-MS. When GC is coupled to a mass spectrometer, the compounds that elute from the GC column are ionized by using electrons (EI, electron ionization) or a chemical reagent (CI, chemical ionization). Charged fragments are focused and accelerated into a mass analyzer: typically a quadrupole mass analyzer. Fragments with different mass to charge ratios will generate different signals, so any compound that produces ions within the mass range of the mass analyzer will be detected. Detection limits of 1-10 ng or even lower values (e.g., 10 pg) can be achieved selecting the appropriate scanning mode.

Sample preparation techniques

Derivatization

Gas chromatography is primarily used for the analysis of thermally stable volatile compounds. However, when dealing with non-volatile samples, chemical reactions can be performed on the sample to increase the volatility of the compounds. Compounds that contain functional groups such as OH, NH, CO 2 H, and SH are difficult to analyze by GC because they are not sufficiently volatile, can be too strongly attracted to the stationary phase or are thermally unstable. Most common derivatization reactions used for GC can be divided into three types:

  1. Silylation.
  2. Acylation.
  3. Alkylation&Esterification.

Samples are derivatized before being analyzed to:

  • Increase volatility and decrease polarity of the compound
  • Reduce thermal degradation
  • Increase sensitivity by incorporating functional groups that lead to higher detector signals
  • Improve separation and reduce tailing

Advantages and disadvantages

GC is the premier analytical technique for the separation of volatile compounds. Several features such as speed of analysis, ease of operation, excellent quantitative results, and moderate costs had helped GC to become one of the most popular techniques worldwide.

Advantages of gc

  • Due to its high efficiency, GC allows the separation of the components of complex mixtures in a reasonable time.
  • Accurate quantitation (usually sharp reproducible peaks are obtained)
  • Mature technique with many applications notes available for users.
  • Multiple detectors with high sensitivity (ppb) are available, which can also be used in series with a mass spectrometer since MS is a non-destructive technique.

Disadvantages of gc

  • Limited to thermally stable and volatile compounds.
  • Most GC detectors are destructive, except for MS.

Gas chromatography versus high performance liquid chromatography (hplc)

Unlike gas chromatography, which is unsuitable for nonvolatile and thermally fragile molecules, liquid chromatography can safely separate a very wide range of organic compounds, from small-molecule drug metabolites to peptides and proteins.

Relative advantages and disadvantages of GC versus HPLC.
GC HPLC
Sample must be volatile or derivatized previous to GC analysis Volatility is not important, however solubility in the mobile phase becomes critical for the analysis.
Most analytes have a molecular weight (MW) below 500 Da (due to volatility issues) There is no upper molecular weight limit as far as the sample can be dissolved in the appropriate mobile phase
Can be coupled to MS. Several mass spectral libraries are available if using electron ionization (e.g., (External Link) ) Methods must be adapted before using an MS detector (non-volatile buffers cannot be used)
Can be coupled to several detectors depending on the application For some detectors the solvent must be an issue. When changing detectors some methods will require prior modification

Bibliography

  • E. F. Barry, Columns for gas chromatography: performance and selection , Wiley-Interscience, Hoboken, NJ (2007)
  • L. S. Ettre, LCGC , 2001, 19 , 120.
  • D. Filmore, Industry Facts&Figures - American Chemical Society Publications . GC: Mature and moving forward. May, 2003.
  • R. L. Grob and E. F. Barry, Modern practice of gas chromatography , 4 th edition, Wiley-Interscience, Hoboken, N.J.(2004)
  • J. V. Hinshaw, LCG , 2013, 31 , 932.
  • A.T. James, Biochem. J. , 1952, 52 , 242.
  • A. T. James and A. J. P. Martin, Biochem. J. , 1952, 50 , 679.
  • A. T. James, A. J. P. Martin, and G. H. Smith, Biochem. J. , 1952, 52 , 238.
  • A. J. P. Martin and R. L. M Synge, Biochem. J. , 1941, 35 , 1358.
  • G. McMahon, Analytical Instrumentation: A Guide to Laboratory, Portable and Miniaturized Instruments , 1 st edition, Wiley, Hoboken, N.J (2007)
  • H. M. McNair, Basic gas chromatography , Wiley, New York (1998)
  • http://www.chromatographyonline.com/
  • Sigma Aldrich Brochure. Derivatization reagents for selective response and detection in complex matrices. 2011,Sigma-Aldrich Co. Available through http://www.sigmaaldrich.com

Get Jobilize Job Search Mobile App in your pocket Now!

Get it on Google Play Download on the App Store Now




Source:  OpenStax, Physical methods in chemistry and nano science. OpenStax CNX. May 05, 2015 Download for free at http://legacy.cnx.org/content/col10699/1.21
Google Play and the Google Play logo are trademarks of Google Inc.

Notification Switch

Would you like to follow the 'Physical methods in chemistry and nano science' conversation and receive update notifications?

Ask