When a radioisotope decays, the daughter product can also be radioactive. Depending upon the relative half-lives of the parent and daughter, several situations can arise: no equilibrium, a transient equilibrium, or a secular equilibrium. This module will not discuss the former two possibilities, as they are off less relevance to this particular discussion.

Secular equilibrium takes place when the half-life of the parent is much longer than the half-life of the daughter. In any arbitrary equilibrium, the ratio of atoms of each can be described as in [link] .

Because the half-life of the parent is much, much greater than the daughter, as the parent decays, the observed amount of activity changes very little.

This can be rearranged to show that the activity of the daughter should equal the activity of the parent.

Once this point is reached, the parent and the daughter are now in secular equilibrium with one another and the ratio of their activities should be fixed. One particularly useful application of this concept, to be discussed in more detail later, is in the analysis of the refinement level of long-lived radioisotopes that are relevant to trafficking.

Detectors

Scintillation detector

A scintillation detector is one of several possible methods for detecting ionizing radiation. Scintillation is the process by which some material, be it a solid, liquid, or gas, emits light in response to incident ionizing radiation. In practice, this is used in the form of a single crystal of sodium iodide that is doped with a small amount of thallium, referred to as NaI(Tl). This crystal is coupled to a photomultiplier tube which converts the small flash of light into an electrical signal through the photoelectric effect. This electrical signal can then be detected by a computer.

Semiconductor detector

A semiconductor accomplishes the same effect as a scintillation detector, conversion of gamma radiation into electrical pulses, except through a different route. In a semiconductor, there is a small energy gap between the valence band of electrons and the conduction band. When a semiconductor is hit with gamma-rays, the energy imparted by the gamma-ray is enough to promote electrons to the conduction band. This change in conductivity can be detected and a signal can be generated correspondingly. Germanium crystals doped with lithium, Ge(Li), and high-purity germanium (HPGe) detectors are among the most common types.

Advantages and disadvantages

Each detector type has its own advantages and disadvantages. The NaI(Tl) detectors are generally inferior to Ge(Li) or HPGe detectors in many respects, but are superior to Ge(Li) or HPGe detectors in cost, ease of use, and durability. Germanium-based detectors generally have much higher resolution than NaI(Tl) detectors. Many small photopeaks are completely undetectable on NaI(Tl) detectors that are plainly visible on germanium detectors. However, Ge(Li) detectors must be kept at cryogenic temperatures for the entirety of their lifetime or else they rapidly because incapable of functioning as a gamma-ray detector. Sodium iodide detectors are much more portable and can even potentially be used in the field because they do not require cryogenic temperatures so long as the photopeak that is being investigated can be resolved from the surrounding peaks.

Get Jobilize Job Search Mobile App in your pocket Now!

Get it on Google Play Download on the App Store Now




Source:  OpenStax, Physical methods in chemistry and nano science. OpenStax CNX. May 05, 2015 Download for free at http://legacy.cnx.org/content/col10699/1.21
Google Play and the Google Play logo are trademarks of Google Inc.

Notification Switch

Would you like to follow the 'Physical methods in chemistry and nano science' conversation and receive update notifications?

Ask